COSMOSIL CN-MS
Features

- Cyanopropyl group bonded stationary phase
- Enables separation of different hydrophobic samples without using gradient
- For separation of natural compounds with different hydrophobicity
Product Information
Column Specifications
Rapid Analysis
Gradient elution is commonly used for the samples containing both polar and non-polar compounds. However, gradient elution may cause reproducibility problem depending on the gradient mixer and pump, and need an equilibration time for each analysis. COSMOSIL 5CN-MS offers rapid analysis and great reproducibility using isocratic elution mode.

| Condition | ||||
|---|---|---|---|---|
| Flow Rate | 1.0 ml/min | Sample | 
 | |
| Temperature | 30°C | |||
| Detection | UV 254 nm | |||
| Small hydrophobicity |  | |||
| Large hydrophobicity |  | |||
Applications
Acetoaminophen

| Condition | ||||
|---|---|---|---|---|
| Column Size | 4.6 mm I.D. x 150 mm | Sample | 
 | |
| Mobile Phase | Methanol/ 50mmol/l KH2PO4 (pH4.7 with NaOH) = 20/80 | |||
| Flow Rate | 1.0 ml/min | |||
| Temperature | 30°C | |||
| Detection | UV 225 nm | |||
Ordering Information
Analytical and Preparative Column (Particle size : 5 µm)
COSMOSIL 5CN-MS Packed Column
(Storage) RT: Room temperature, A: Cool and dark, R: Refrigerator, F: Freeze