COSMOSIL C18-AR-II
Features
- Features strong acid resistance
- Suitable for acid compounds and peptides
Product Information
Column Specifications
Acid Resistance
The COSMOSIL 5C18-AR-II packed column features a polymeric type of C18 reversed phase material. The acidic resistance of COSMOSIL 5C18-AR-II is much improved compared with commercially available monomeric type octadecyl stationary phases. It retains high performance even in case of acidic mobile phases commonly used to separate acidic compounds and peptides.
Figure.
Decomposition test in 0.1% Trifluoroacetic Acid solution at 60°C .
Capacity factor (k') = Naphthalene,
Mobile phase: 70% methanol
Applications
Peptides
Condition | |
---|---|
Column Size | 4.6 mm I.D. x 150 mm |
Mobile Phase | A; 0.05%TFA-H2O B; 0.05%TFA-Acetonitrile B conc. 10→40% 20min Linear gradient |
Flow Rate | 1.0 ml/min |
Temperature | 30°C |
Detection | UV220nm |
Sample |
|
Organic Acids
Condition | |
---|---|
Column Size | 4.6 mm I.D. x 150 mm |
Mobile Phase | Methanol/ 20mmol/l Phosphoric Acid = 20/80 |
Flow Rate | 1.0 ml/min |
Temperature | 30°C |
Detection | UV254nm, 0.5AUFS |
Sample |
|
Salicylic Acid Esters
Condition | |
---|---|
Column Size | 4.6 mm I.D. x 150 mm |
Mobile Phase | Acetonitrile/ H2O = 70/30 |
Flow Rate | 1.0 ml/min |
Temperature | 30°C |
Detection | UV254nm, 0.2AUFS |
Sample |
|
Parabens
Condition | |
---|---|
Column Size | 4.6 mm I.D. x 150 mm |
Mobile Phase | Acetonitrile/ H2O = 50/50 |
Flow Rate | 1.0 ml/min |
Temperature | 30°C |
Detection | UV254nm, 0.12AUFS |
Sample |
|
LC/MS Applications
Usage information
Column | 5 µm C18-AR-II | |||||||
---|---|---|---|---|---|---|---|---|
I.D. (mm) | 1.0 | 2.0 | 3.0 | 4.6 | 6.0 | 10.0 | 20.0 | 28.0 |
Washing method | 1. Remove buffer, salts and/or acid from column: wash for 10-15 min. using the mobile phase last used, without buffer, salts or acid. For example, if your mobile phase was 50:50 methanol/20 mmol/l phosphate buffer, wash with 50:50 methanol/water. 2. (If you used a buffer, always do step 1 first!) Remove adsorbed compounds and fix unstable baselines by washing with methanol and/or THF. |
|||||||
Storage conditions | Short-term (up to a week): Remove buffer, salts and/or acid from column: wash for 10-15 min. using the mobile phase last used, without buffer, salts or acid. For example, if your mobile phase was 50:50 methanol/20 mmol/l phosphate buffer, wash with 50:50 methanol/water. Long-term: Remove buffer, salts and/or acid from column: wash for 10-15 min. using the mobile phase last used, without buffer, salts or acid. Then, replace the solvent with 70% methanol or acetonitrile : 30% water. In either case, tightly plug the column, and store in a cool and dry place. |
|||||||
Usable pH range | pH1.5~pH7.5 (pH 2-7.5 recommended) |
|||||||
Max. pressure | 20MPa | 15MPa | ||||||
Temperature range | The maximum usable temperature is 60°C. However. for regular use, please use at a constant temperature between 20°C and 50°C. | |||||||
Usable solvents | Any solvent that will not dissolve the silica gel (such as alkaline solutions) or cleave the stationary phase (such as very acidic solutions) is usable. Note: For solvents with high viscosity, please watch the system pressure and keep it below 20 MPa. Also, please wash the column after using acidic mobile phases or solvents with high freezing points. |
|||||||
Other instructions | - Buffer concentration is usually sufficient at 0.005 - 0.02 mol/L. - Always filter mobile phases using a 0.5 µm or finer filter before use. - Reproducibility may worsen when using a mobile phase that is more than 90% water. |
Downloads
Brochure
COSMOSIL 5C18-AR-II (659 KB)
References
Reference list
Ordering Information
Analytical Column (Particle Size : 5 µm)
COSMOSIL 5C18-AR-II
4.6 mm I.D. x 10 mm Cartrige contains 3 PKG. Cartridge Holder (Product number: 38009-79) is required.
Preparative Column (Particle Size : 15 µm)
COSMOSIL 15C18-AR-II
For more information on 15C18-AR-II
Fast LC Column (Particle Size : 3µm)
COSMOSIL 3C18-AR-II
Product Name | Size | Product No. | Price |
---|---|---|---|
COSMOSIL 3C18-AR-II Packed Column | |||
2.0 mm l.D. x 50 mm | 05478-91 | ||
4.6 mm l.D. x 10 mm | 38068-41 | ||
4.6 mm l.D. x 50 mm | 38069-31 | ||
4.6 mm l.D. x 100 mm | 38070-91 |
- For more information on COSMOSIL 3.0 mm I.D. columns
- Please refer to Inner Diameter of Column(scale down and scale up) (PDF 344 KB)
(Storage) RT: Room temperature, A: Cool and dark, R: Refrigerator, F: Freeze