Products

COSMOSIL πNAP

Features

Column Image

  • Naphthalene bonded stationary phase
  • Enhanced π-π interactions
  • Improved selectivity for structural isomers

For information on COSMOSIL 2.5πNAP (Ultra-High Performance Column)

Product Information

Column Specifications

Column Specifications

Comparison of π-π interactions

COSMOSIL πNAP shows stronger π-π interactions than phenyl columns. Its two fused aromatic rings retain nitrobenzene with more π electrons stronger than phenyl columns.

pinap

Applications

Tolunitrile

Tolunitrile

Condition
Column Size 4.6 mm I.D. x 150 mm Tolunitrile
Flow Rate 1.0ml/min
Temperature 30°C
Detection UV 254 nm
Sample
  • o-Tolunitrile 2.0 µg
  • m-Tolunitrile 2.0 µg
  • p-Tolunitrile 1.0 µg

Adrenal Cortical Hormone

Adrenal cortical hormone

Condition
Column Size 4.6 mm I.D. x 150 mm Adrenal cortical hormone
Flow Rate 1.0ml/min
Temperature 30°C
Detection UV 254 nm
Sample
  1. Prednisolone 0.33 µg
  2. Hydrocortisone 0.33 µg
  3. Cortisone 0.33 µg

Berberine

Berberine

Condition
Column Size 4.6 mm I.D. x 150 mm Berberine
Mobile Phase Methanol : 20 mmol/l Phosphate Buffer (pH2.5) =
5C18-MS-II 40 : 60
5PE-MS 60 : 40
πNAP 80 : 20
Flow Rate 1.0ml/min
Temperature 30°C
Detection UV 254 nm
Sample
  1. Palmatine chloride 0.2 µg
  2. Berberine Hydrochoride 0.2 µg

Isoflavone

Isoflavone

Condition
Column Size 4.6 mm I.D. x 150 mm Flow rate 1.0ml/min
Mobile Phase A Acetonitrile / 20mmol/l Phosphate Buffer (pH 2.5)=10/90
B Acetonitrile / 20mmol/l Phosphate Buffer (pH 2.5)=50/50
B Conc. 0 → 100% 20min Linear Gradient
Temperature 30°C
Detection UV 260 nm
Sample
  1. Daidzin 0.05 µg
  2. Glycitin 0.05 µg
  3. Genistin 0.05 µg
  4. Daidzein 0.05 µg
  5. Glycitein 0.075 µg
  6. Genistein 0.025 µg

See more COSMOSIL Applications

Usage information

Column 5 µm πNAP
I.D. (mm) 2.0 3.0 4.6 10.0 20.0 28.0
Washing method 1. Remove buffer, salts and/or acid from column: wash for 10-15 min. using the mobile phase last used, without buffer, salts or acid. For example, if your mobile phase was 50:50 methanol/20 mmol/l phosphate buffer, wash with 50:50 methanol/water.
2. (If you used a buffer, always do step 1 first!) Remove adsorbed compounds and fix unstable baselines by washing with methanol and/or THF.
Storage conditions Short-term (up to a week):
Remove buffer, salts and/or acid from column: wash for 10-15 min. using the mobile phase last used, without buffer, salts or acid. For example, if your mobile phase was 50:50 methanol/20 mmol/l phosphate buffer, wash with 50:50 methanol/water.

Long-term:
Remove buffer, salts and/or acid from column: wash for 10-15 min. using the mobile phase last used, without buffer, salts or acid. Then, replace the solvent with 70% methanol or acetonitrile : 30% water.

In either case, tightly plug the column, and store in a cool and dry place.
Usable pH range pH2~pH7.5
Max. pressure 20MPa 15MPa
Temperature range The maximum usable temperature is 60°C. However. for regular use, please use at a constant temperature between 20°C and 50°C.
Usable solvents Any solvent that will not dissolve the silica gel (such as alkaline solutions) or cleave the stationary phase (such as very acidic solutions) is usable.
Note: For solvents with high viscosity, please watch the system pressure and keep it below 20 MPa. Also, please wash the column after using acidic mobile phases or solvents with high freezing points.
Other instructions - Buffer concentration is usually sufficient at 0.005 - 0.02 mol/L.
- Always filter mobile phases using a 0.5 µm or finer filter before use.
- Acetonitrile is not recommended as a mobile phase.
- Baseline drift may occur.

Downloads

Brochures

Posters

  • Evaluation of Pentabromobenzyl Stationary Phase for Separation of Polar Compounds by Reversed-Phase Chromatography
    (Poster from HPLC 2015, Geneva)
    pdf (370 KB)

References

Reference list

Ordering Information

Analytical / Preparative Column (Particle Size : 5 µm)
COSMOSIL πNAP

Product Name Size Product No. Price
COSMOSIL πNAP Guard Column
4.6 mm I.D. x 10 mm 08082-71 e-Nacalai
10 mm I.D. x 20 mm 08087-21
20 mm I.D. x 20 mm 08090-61
20 mm I.D. x 50 mm 08091-51
28 mm I.D. x 50 mm 08094-21
COSMOSIL πNAP Packed Column
1.0 mm I.D. x150 mm 08076-61 e-Nacalai
1.0 mm I.D. x 250 mm 08077-51
2.0 mm I.D. x 150 mm 08078-41
2.0 mm I.D. x 250 mm 08079-31
3.0 mm I.D. x 150 mm 08080-91
3.0 mm I.D. x 250 mm 08081-81
4.6 mm I.D. x 50 mm 08083-61
4.6 mm I.D. x 100 mm 08084-51
4.6 mm I.D. x 150 mm 08085-41
4.6 mm I.D. x 250 mm 08086-31
10 mm I.D. x 150 mm 08088-11
10 mm I.D. x 250 mm 08089-01
20 mm I.D. x 150 mm 08092-41
20 mm I.D. x 250 mm 08093-31
28 mm I.D. x 250 mm 08095-11

Ultra-High Performance Column (Particle Size : 2.5 µm) 
COSMOSIL 2.5πNAP

Product Name Size Product No. Price
COSMOSIL 2.5πNAP Packed Column
2.0 mm l.D. x 50 mm 06062-91 e-Nacalai
2.0 mm l.D. x 75 mm 06051-31
2.0 mm l.D. x 100 mm 06052-21
3.0 mm l.D. x 50 mm 06054-01
3.0 mm l.D. x 75 mm 06055-91
3.0 mm l.D. x 100 mm 06057-71

For information on COSMOSIL 2.5πNAP (Ultra-High Performance Column)

(Storage) RT: Room temperature, A: Cool and dark, R: Refrigerator, F: Freeze